Comparison of Plasma With Whole Blood Prothrombin Time and Fibrinogen on the Same Instrument
Abstract
We compared plasma with whole blood (WB) international normalized ratio (INR) and fibrinogen using the same instrument and reagents. WBINRs were 50% higher than plasma INRs. After increasing the WB sample volume 40% and adjusting the International Sensitivity Index, WBINRs were similar to plasma INRs [adjusted WBINR = 0.99(plasma INR) – 0.02; r2 = 0.98; n = 155], but the average difference in WB vs plasma INR was 4-fold higher than duplicate plasma INRs. Variation in hematocrit was a major determinant of the accuracy of the WBINR, with increased error at high INRs. The WB fibrinogen assay was highly dependent on the sample hematocrit (r2 = 0.83), even after the sample volume was adjusted. Accurate WB fibrinogen measurements required a mathematical hematocrit correction. We conclude that WBINR and fibrinogen assays can be performed on point-of-care or automated analyzers, but sample volume must be adjusted to account for hematocrit. Accuracy is limited by variations in hematocrit with worsening accuracy for samples with high INRs or low fibrinogen levels.
Materials and Methods
Whole blood (WB) prothrombin time (PT) and international normalized ratio (INR) testing is used for outpatient monitoring of warfarin therapy and to reduce turnaround time. WB testing typically uses finger-stick or citrated WB specimens, whereas plasma INR, the reference method, uses citrated plasma specimens. Studies comparing WBINR with plasma INR generally show correlations of r2 = 0.70 to 0.96,1–6 but WBINR tends to be biased high, with the bias becoming worse as the INR rises.1,6–12
In contrast with WBINR testing, direct WB testing for individual coagulation factors is not widely used or widely available. Indirect WB fibrinogen viscoelastic assays have been devised based on the fact that the maximal amplitude is primarily a function of the platelet count and fibrinogen concentration in these assays.13,14 The maximum amplitude is most closely related to the fibrinogen concentration in the presence of specific fibrinogen activators (such as reptilase) or glycoprotein IIb/IIIa receptor antagonists that block platelet binding to fibrin.15–17 However, correlation between viscoelastic fibrinogen and plasma fibrinogen measurements is relatively weak (r2 < 0.61) with limited positive predictive value for low fibrinogen levels (<55%).15–17
Potential sources of inaccuracy between plasma and WB methods include thromboplastin sensitivity,12,18 differences in clot detection mechanisms,19 citrate,11,20 coagulation factor levels,9 technical expertise,21,22 intrinsic imprecision of WB methods, and hematocrit values.23 When point-of-care instruments are compared with automated analyzers, it is difficult to determine which source of inaccuracy between WB and plasma assays is most important because multiple differences are present in the same comparison. The purpose of our study was to evaluate the effect of hematocrit on WB coagulation testing, while eliminating other sources of variation. Hematocrit error applies to all WB assays and is, therefore, an important variable to understand. Our study compares WB and plasma PT and fibrinogen measurements using the same instruments, reagents, and technical expertise. This design allowed us to control for many preanalytic and analytic variables and determine the inherent performance limitations of WBINR and WB fibrinogen measurements. It also allowed us to assess the feasibility of performing WB coagulation testing on an automated coagulation analyzer.
The study was approved by the University of Washington Human Subjects Review Committee (Seattle). Venous blood samples were anticoagulated by adding 2.7 mL of blood to 0.3 mL of 0.109 mol/L citrate. All samples were gently mixed manually and then split. The aliquot for plasma studies was centrifuged at 3,600g for 2 minutes at room temperature and analyzed.
Results
Subjects and Samples
All PT and INR measurements were made on Diagnostica Stago analyzers (STA-R Evolution and STA-Compact, Diagnostica Stago, Parsippany, NJ) using Neoplastine CI Plus reagents (Diagnostica Stago). The International Sensitivity Index (ISI) and geometric mean normal for the PT reagent were 1.25 and 13.1 seconds, respectively. For the plasma PT assay, a 50-μL sample of citrated plasma was incubated for 4 minutes at 37°C, followed by addition of 100 μL of STA prothrombin reagent. The incubation period was used to bring the plasma sample to 37°C before adding the PT reagent. Although the Stago instrument used a standard incubation time of 4 minutes, only 1 minute was actually required to bring the sample up to temperature. For WBPT measurements, samples were loaded singly instead of in batches, and the PT incubation time was decreased from 4 minutes to 1 minute. These adjustments were done to allow assay completion before significant settling out of the cellular components. Correlation studies done on samples measured with 1- and 4-minute incubation times showed no significant difference in results.
PT and INR Measurements
Owing to the presence of cells in the WB sample, an equal sample volume of WB delivers less than the expected volume of plasma to the assay. It was not possible or practical to adjust the volume for each WB sample to deliver 50 μL of plasma to the PT assay. For adjusted WBPT (AWBPT) measurements, the citrate WB sample volume was increased by 40% to 70 μL to deliver approximately 50 μL of citrate plasma volume to the assay when the hematocrit value was near 30% (0.30; average for patients having stat coagulation testing ordered). In the AWBPT assay, well-mixed WB samples were loaded singly instead of in batches, and a 70-μL sample of citrated WB was incubated for 1 minute, followed by addition of 100 μL of prothrombin reagent.
We evaluated the effect of hematocrit on the PT in samples at a range of INRs by spiking split plasma samples with increasing amounts of WB cellular components from the same subject.
Open in new tab
Effect of Hematocrit Value on PT
Discussion
Plasma Fibrinogen Measurements
With the same 50-μL sample volume for plasma and WBINR assays and the original plasma ISI of 1.25, WBINR values were on average more than 50% higher than the plasma INR (Figure 1). The AWBINR had 2 modifications. First, the AWBPT method used the increased 70-μL WB sample volume, and second, the ISI was adjusted to 1.30 to provide the best fit between WB and plasma INR values. On average, there was little bias between plasma INR and AWBINR results but, again, significant variability. Duplicate plasma INRs showed an average ± SD difference of 0.02 ± 0.05. Duplicate AWBINRs showed an average ± SD difference of 0.01 ± 0.1. For plasma INR vs AWBINR, the average ± SD difference was overall −0.04 ± 0.2, which increased to −0.11 ± 0.4 for INRs greater than 2.
Figure 2 shows a comparison of plasma INR vs AWBINR at different hematocrit values in the same sample. For samples with plasma INRs of 1.24 and 2.17, the mean AWBINR was not significantly different from the plasma INR, and there was no correlation between AWBINR and hematocrit values. For plasma INRs of 3.04 and 4.69, the mean AWBINR was significantly less than the plasma INR, but there was no correlation between AWBINR and hematocrit values. For a plasma INR of 5.7, the AWBINR was significantly correlated with hematocrit (r2 = 0.98; P = .001).
We studied the concordance of the AWBINR vs plasma INR using INR-defined cohorts Table 1. The groups were INR less than 2, 2 to 4, and more than 4. The overall concordance was 97% between the 2 assays for the 154 samples studied, but it worsened with increased INR. At INRs less than 2, the concordance was 98%, and at INRs more than 4, the concordance was 88%.
The AWBFIB level was on average only 2% to 4% lower than the plasma fibrinogen level but showed substantial variation Figure 3. For plasma fibrinogen levels less than or equal to 150 mg/dL (1.5 g/L), the average ± SD difference between plasma and AWBFIB was 6 ± 20 mg/dL (0.06–0.2 g/L). For plasma fibrinogen levels greater than 150 mg/dL (1.5 g/L), the average ± SD difference between plasma and AWBFIB was −23 ± 56 mg/dL (−0.23 ± 0.56 g/L). Even though the AWBFIB sample volume was adjusted to increase the sample plasma volume to approximately the same level as the plasma assay, hematocrit still had a major effect on the fibrinogen level measured in WB Figure 4 (r2 = 0.80). If the hematocrit value was known, a CWBFIB level could be derived that substantially reduced the variation between WB and plasma fibrinogen measurements (Figure 3). The average ± SD difference between CWBFIB and plasma fibrinogen levels was 0.5 ± 21 mg/dL (0.005–0.21 g/L).
Hematocrit was an important determinant of the accuracy of WBINR and fibrinogen measurements. Point-of-care and viscoelastic coagulation assays are designed to correct for an average hematocrit value in patient samples to address this issue. However, our data demonstrate that a single assumed sample hematocrit value or a single arithmetic correction is not sufficient. The impact of hematocrit values remains significant, especially in samples with hematocrit values more than 10% from the corrected-for hematocrit value and for samples with high INRs. Hence, the sickest patients with high or low hematocrit values (eg, patients with anemia, substantial blood loss, polycythemia, or dehydration) are most at risk for a clinically significant error, particularly if they have INRs more than 3.0. A majority of studies comparing WB with plasma INR testing demonstrated biases that were more pronounced at higher INRs,1,2,7–12,24 with the magnitude and direction of discrepancies and biases varying significantly even between sites in a single study.12,18,24 Our results suggest that an important determinant of the magnitude and direction of the bias is how different the hematocrit value of a given sample is from the corrected-for hematocrit value of the WB testing platform in question.
van den Besselaar et al23 investigated the impact of hematocrit values on differences between WB and plasma INRs and concluded that the hematocrit value had a statistically but not clinically significant effect on the INR in the hematocrit range from 37% to 51% (0.37–0.51). However, their data showed that samples with hematocrit values less than 35% (0.35) would result in WB vs plasma INR differences more than 0.65 INR based on hematocrit values alone and that the discrepancy due to hematocrit increases as samples get further away from their corrected-for hematocrit value of 40% (0.40).
Our data, in addition to showing similar trends, demonstrate that the effect of hematocrit values on INR measurements is amplified at higher INRs. The relationship of INR to plasma factor level is exponential.25 Hence, low INRs are relatively insensitive to factor level changes and, thus, plasma sample volume differences. The INR becomes increasingly sensitive to factor level and, thus, sample plasma volume differences as factor levels decrease. This exponential relationship between the PT/INR and factor levels precludes the use of a simple hematocrit-based arithmetic correction to improve WB coagulation testing accuracy and implies an analytic limitation for WBINR testing on current platforms. An 18-month–28-site evaluation of a point-of-care (WB) coagulation instrument by Maddox et al11 required graduated INR-based correction factors in order to improve agreement between the WB and plasma results. The study by Maddox et al,11 like other WB-vs-plasma comparison studies, had large biases that were more pronounced at higher INRs and found that a single simple arithmetic correction was not sufficient to correct the bias across the range of INRs. While WBINR testing may be sufficiently accurate below an INR of 4, higher WBINR values need to be repeated by a plasma method. Another explanation for discrepancies between plasma and WB methods is that WB testing is inherently less precise than plasma-based methods. Our results show that WBINR and fibrinogen are about 4-fold less precise than the plasma-based method.
Statistical Analysis
Like INR, the hematocrit value is an important determinant of the accuracy of WB fibrinogen assays. Our study shows that although the accuracy and usefulness of WB fibrinogen assays is improved by accounting for the impact of hematocrit values, a single arithmetic correction based on a population hematocrit value is not sufficient. Even after applying the population-hematocrit–based correction, we still had differences of up to 200 mg/dL (2.0 g/L) between plasma and WB fibrinogen levels for a single specimen with a study design that used identical instruments and reagents. Samples from patients with values 10% or more from the corrected-for hematocrit value would require an individual correction to have clinically reliable accuracy, which may limit the turnaround time benefit of WB viscoelastic fibrinogen testing.
WB vs Plasma PT and INR
WB viscoelastic fibrinogen assays use the relationship between the clot strength (reported alternatively as MA, maximal clot firmness or shear elasticity) and fibrinogen concentration as a basis for determining specimen fibrinogen concentration. In keeping with our findings of the impact of the hematocrit value, the relationship between clot strength and fibrinogen concentration is stronger for plasma specimens (r2 > 0.94)16,26 than for WB specimens (r2 < 0.605).15,16,27
A and B, Deming regression analysis and difference plots for the adjusted whole blood fibrinogen (AWBFIB) vs plasma fibrinogen (FIB) (n = 158) equation for the line AWBFIB = 0.98x – 11.83; r2 = 0.929. C and D, Deming regression analysis and difference plots for the corrected whole blood fibrinogen (CWBFIB) vs plasma FIB (n = 126) equation for the line CWBFIB = 1.01x – 4.76; r2 = 0.984. CWBFIB values were adjusted mathematically for differences in hematocrit values.
The only commercially available WB fibrinogen assay we found was the Hemochron Whole Blood Coagulation System Fibrinogen assay (International Technidyne, Piscataway, NJ). This is a Clauss-type fibrinogen assay that uses WB rather than plasma. This WB fibrinogen assay showed only a moderate correlation (r2 = 0.88) with the plasma fibrinogen assay (package insert), although the validation study was performed using WB specimens from a single donor, which reduces the impact of variations in hematocrit values on the true accuracy of the assay.
With regard to WB testing on a standard coagulation analyzer, we found that several modifications were necessary. First, because the coagulation analyzer was not designed to mix samples as hematology analyzers are, only a few WB samples could be analyzed at a time before cell settling occurred. Second, if WB testing is contemplated, adjustment of the sample volume is needed to account for the average hematocrit value of patient samples that will be tested. This may vary depending on whether the population is an out-patient population with a near normal hematocrit value or a postoperative or trauma population that may have a lower hematocrit value. Third, as described, differences in hematocrit values produced variations in the measured PT/INR and fibrinogen values between WB and plasma results. This variation was not so severe as to preclude use of WBPT/INR, particularly for INRs near or below the therapeutic range, but it may be unacceptable for WB fibrinogen measurement unless a hematocrit correction is used. This requires that a hematocrit value be obtained with each WB fibrinogen value, which may be acceptable for certain stat panels, but may limit use of WB fibrinogen testing when the hematocrit value is not available. If these modifications are incorporated, WB testing is sufficiently reliable for clinical testing.